Agilent 990 Micro GC Micro Gasifier User Manual
21
2
Installation and Use
Operation
Operation
1
After all the connections are made, allow 15 minutes stabilization of the temperature.
2
Connect the sample bomb to the Micro Gasifier. In case of a LPG, ensure that the liquid phase
is flowing towards the Micro Gasifier.
3
Open the sample bomb, and check for leaks.
4
Open the VENT needle valve, and allow to flush for one minute, then check the sample vent
flow at the VENT-OUTLET.
5
Open the SAMPLE needle valve, and adjust a flow though the Micro GC of 10 mL/min.
6
Close the VENT needle valve to a flow of 10 mL/min minimum.
7
Start the Micro GC run or sequence.
Monitoring the various flows by means of a soap bubble meter, or equivalent, will help setting the
optimal conditions.
In case of a high-pressure gas, the same procedure applies.
Figure 1.
Product diagram
1/16"
Connection
Connection to GC
Isolated housing
1/16"
Connection
1/16"
Connection
Vent
Pressure
regulator/
Evaporator
12 Volt
for heating
Liquid
in
Содержание 990
Страница 1: ...Agilent 990 Micro Gas Chromatograph Micro Gasifier User Manual ...
Страница 4: ...4 Agilent 990 Micro GC Micro Gasifier User Manual ...
Страница 10: ...1 Introduction Disposal Instructions 10 Agilent 990 Micro GC Micro Gasifier User Manual ...
Страница 23: ...Agilent 990 Micro GC Micro Gasifier User Manual 23 ...
Страница 24: ...www agilent com Agilent Technologies Inc 2019 First edition August 2019 G3588 90025 G3588 90025 ...