705
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_cctphase.fm
Code Channel Time/Phase Error Measurement Troubleshooting
Code Channel Time/Phase Error Measurement Troubleshooting
Possible Setup Issues
IMPORTANT
If the Access Probe Power measurement is currently selected and the Trigger Arm setting is
Continuous, all other measurements are disabled. Use the
command to set the trigger arm state to single and retry this measurement.
During remote operation of this measurement, you should configure the Trigger Arm to Single. See
. If single triggering is used during manual operation, press the START SINGLE key to
trigger each measurement.
The input signal must conform to
“Input Signal Requirements” on page 51
.
Digital average power measurement results are affected by the Amplitude Offset value. Use the
command to determine offset values.
The test set receiver’s expected signal amplitude and channel number (or frequency) must be set up correctly
for accurate measurements. This is normally done automatically (see
) but can
also be done manually (see
“Receiver Control” on page 194
) .
Interpreting Integrity Indicator values
See
“Integrity Indicator” on page 222
.
If over range (5) is returned the input power has exceeded the test set’s internal sampler maximum value
during some part of the sampling or the input power has exceeded the calibrated range of the test set’s power
detector.
If under range (6) is returned the measurement result is below the expected input power level. Under range is
also indicated if the input power is below the calibrated range of the test set’s power detector.
If the signal has both over range and under range conditions, only the over range (5) will be indicated.
Related Topics
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...