372
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_rfanalyzer.fm
RFANalyzer
RFANalyzer:CONTrol:AUTO
RFANalyzer:EXPected:POWer[:SELected]
Function
Sets/queries the test set’s receiver control. The measuring receiver is under the control of the
test set’s base station emulator (auto) or under the control of the user (manual). see
Setting the manual band will change the receiver control to manual. see
“RFANalyzer:MANual:BAND” on page 375
Setting the manual channel in the band that is currently active will change the receiver control
to manual. see
“RFANalyzer:MANual:CHANnel[:SELected]” on page 375
Setting the manual frequency will change the receiver control to manual. see
“RFANalyzer:MANual:FREQuency” on page 378
Setting the broadcast band will change the receiver control to auto. See
Setting
Manual = 0|OFF , Auto = 1|ON (default ON)
Query
0|1
*RST setting
0|on
Programming Example
OUTPUT 714;”RFANALYZER:CONTROL:CONTROL:AUTO 0” !Sets receiver control to manual.
Function
Sets/queries the power level in DBM that the mobile station is expected to transmit for the
selected band. The units DBM are optional. The test set will set up its input signal path to
measure this power level when a user is in manual control. See
Setting
Range: -60 to +53 dBm, after the Amplitude Offset (optional) has been factored in. see
“Measurement Related Configuration” on page 563
Resolution: .01 DBM
Query
Range: -60 to +53 dBm, after the Amplitude Offset (optional) has been factored in. see
“Measurement Related Configuration” on page 563
Resolution: .01 DBM
*RST setting
Sets the receiver control to auto
Band: PGSM
Expected Power: +13DBM
Programming Example
OUTPUT 714;”RFANALYZER:EXPECTED:POWER:SELECTED 10” !Sets the test set expected
!input level to 10 DBM.