10

Setting the Channel and Protocol

Access the CDMA CALL CONTROL screen, see
Figure 11.

1. Set the 

RF Chan Std

field (item 3) to the appro-

priate standard.

1a. For example, for U.S. PCS frequencies, set

this field to 

US PCS.

1b. For U.S. cellular frequencies, set this field

to 

MS AMPS.

1c. For CDMA phones for use in Japan, choose

Japan CDMA.

The 

RF Channel

field is auto-

matically updated with the starting chan-
nel within the given standard. 

2. Enter the appropriate channel value in the 

RF

Channel

field (item 1). The manufacturer or

service provider programs the phone to the
appropriate RF channels. If the RF channel is
not known or incorrectly entered into the test
set, the phone will start to scan all the possible
channels in the phone’s channel list. If an out-
side base station of higher power is received,
the phone will lock onto it and discontinue its
search. For some phones, this is a long process,
therefore, it is recommended that the appropri-
ate channel value is known and entered into
the 

RF Channel

field.

3. Enter the appropriate protocol for the selected

channel standard into the 

Protocol

field (item 2).

For example: 

3a. For U.S. CDMA PCS frequencies, select

JSTD-008.

3b. For U.S. CDMA cellular frequencies, select

IS-95A

or 

TSB-74.

3c. For Japanese CDMA, select 

ARIB T53.

Figure 11. CDMA CALL CONTROL Screen

Содержание 8924C

Страница 1: ...hod 18 Using the Agilent 8924C Test Set 18 Appendix 4 Trigger Signal Description 19 Making a Trigger Cable 20 Trigger Signal Description 19 Reference 19 Open Loop Power Control Time Response Test This...

Страница 2: ...0 dBm base station See Figure 2 2 The test set power is increased by a step of 20 dB to a power level of 40 dBm 3 The test set power is decreased by a step of 20 dB to a power level of 60 dBm 4 The te...

Страница 3: ...easurements Option 160 CDMA firmware for use with option 151 OTHER SUGGESTED OPTIONS Option 041 GPIB and Parallel Interface Option 105 Time Gated Spectrum Analysis To determine which spectrum analyzer...

Страница 4: ...et for cellular frequencies or to the RF IN OUT port on the PCS interface for PCS frequencies 4 Connect the trigger cable from the AUX CON TROL port 15 pin D connector on the back of the test set to t...

Страница 5: ...5 Figure 4 Setup for PCS Band Phones...

Страница 6: ...onfiguring the Test Set 3 Set the RF power increment to 20 dB on the test set as described in Section 6 Setting Up a 20 dB Increment on the Test Set 4 Calibrate channel power on the test set by access...

Страница 7: ...pt 2 9b Press the test set s CALL key to connect a call in data loopback mode at full rate The phone will usually display a message indicating LOOP MODE 10 Once a call is connected go to the CDMA TRAN...

Страница 8: ...ectrum analyzer 16 Step decrease Sctr A Pwr field to 80 dBm The new time response will be displayed on the spectrum analyzer 17 Step increase Sctr A Pwr field to 60 dBm The new time response will be d...

Страница 9: ...order to activate the RF In Out field item 2 2 The RF In Out field should be set to the attenua tion value determined by the methods described in Appendix 3 Determining Cable Connection Losses This wi...

Страница 10: ...phone to the appropriate RF channels If the RF channel is not known or incorrectly entered into the test set the phone will start to scan all the possible channels in the phone s channel list If an o...

Страница 11: ...elect Chan Power item 5 For this test the channel power meter is used because it gives a channel power reading specific to the reverse channel bandwidth that is occupied by the phone 6 The Power Meas...

Страница 12: ...tely determine the delay between the trigger sig nal and the actual power step at the test set s out put Inaccurate delay times will cause the test to start before the power level from the test set ha...

Страница 13: ...s the EXTERNAL softkey NOTE The spectrum analyzer is now ready to measure the time delay between the E8285A trigger signal going high and the E8285A RF output level changing 20 dB 10 Change the test s...

Страница 14: ...channel 3 Press the MS EXT ATTEN softkey on sub menu 1 of 3 enter the value for the attenuation meas ured from the power splitter and the cables connecting the test set to the spectrum analyzer This i...

Страница 15: ...the Time Response Function 1 Press the More 1 of 3 softkey on the CDMA ANA LYZER main menu to go to menu 2 of 3 see Figure 16 2 Press the Time Response softkey to access the time response test functio...

Страница 16: ...r an accurate delay time in order to obtain accurate test results The test may fail frequently if the trigger does not occur at the time of power transitions 3d Press the Previous Menu softkey to retu...

Страница 17: ...nstant The open loop power control response for cellular and PCS frequencies are described by the following equations Cellular Equation Receive Power Forward Link Transmit Power Reverse Link Nominal P...

Страница 18: ...trol mode but with an error of 1 dB at best This can cause a problem when the base station is transmitting near the noise floor Since there is no form of error correction associated with the punc ture...

Страница 19: ...two cables an external power meter can be used Listed below are the steps you need to take to find your RF power compensation 1 Zero the power meter 2 Connect the power meter directly to the test set...

Страница 20: ...the mobile s open loop power control notices that it is transmit ting the reverse signal at too high a power and starts to lower its signal strength This variation is undesirable for the open loop tim...

Страница 21: ...BNC ground lug 4 Shrink wrap the D connector so it will not contact any pins 5 Shrink wrap the conductor and soldered pins and install the hood on the D connector Trigger Signal Description Approximat...

Страница 22: ...Your Advantage means that Agilent offers a wide range of additional expert test and measurement services which you can purchase according to your unique technical and business needs Solve problems ef...

Отзывы: