13 Maintaining the NPD
To Maintain the NPD Collector, Ceramic Insulators, and Jet
220
Maintaining Your GC
To Maintain the NPD Collector, Ceramic Insulators,
and Jet
When replacing the jet, always install a new collector, ceramic insulators, and metal C-rings.
When replacing the collector, Agilent recommends replacing the ceramic insulators and metal
C-rings.
WARNING
The insulation around the inlets, detectors, valve box, and the insulation cups is made of
refractory ceramic fibers. To avoid inhaling fiber particles, we recommend the following
safety procedures: ventilate your work area; wear long sleeves, gloves, safety glasses, and a
disposable dust/mist respirator; dispose of insulation in a sealed plastic bag; wash your
hands with mild soap and cold water after handling the insulation.
1
Gather the following:
•
NPD ceramic insulator kit, see
“Consumables and Parts for the NPD”
•
Collector
•
Cap for the bead
•
T-10 and T-20 Torx screwdrivers
•
Tweezers
•
Cotton swab
•
Solvent
•
Methanol
•
Jet
•
Lint-free gloves
•
Compressed, filtered dry air or nitrogen
CAUTION
The bead is delicate. Be careful not to break or crack the bead. When performing
maintenance on the NPD, avoid touching the bead with your fingers, and prevent it from
coming in contact with other surfaces.
2
Turn off the bead and disable
Adjust Offset
.
3
Check and note the NPD leakage current for reference. See
4
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready to continue (all
zones cooled to safe handling temperatures).
WARNING
Be careful! The oven or detector fittings may be hot enough to cause burns.
5
Remove the bead. See
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...