6-22
Performance Tests – Option 011
7. Receiver Magnitude Frequency Response
3
Press
PRESET
,
MARKER
,
MEAS
,
INPUT PORTS
,
R
,
MENU
,
POWER
,
6
,
x1
.
4
Press
CW FREQ
,
300
,
k/m
. Set the calibration factor on the power meter for this
CW frequency.
5
Write the power meter reading on the “Performance Test Record.”
6
Write the marker reading, which appears on the analyzer display, in the “R
Input Power” column of the “Performance Test Record.”
7
through
for the other CW frequencies listed on the
“Performance Test Record.”
Input A Magnitude Frequency Response
8
Connect the equipment as shown in
.
Figure 6-10. Magnitude Frequency Response Test Setup (Receiver Input A)
9
Press
MEAS
,
INPUT PORTS
,
A
,
MENU
to measure the power at the receiver input A
channel.
10
Press
CW FREQ
,
300
,
k/m
.
11
Write the marker reading, which appears on the analyzer display, in the
“A Input Power” column of the “Performance Test Record.”
12
for the other CW frequencies listed on the
“Performance Test Record.”