5-7
Automated Verification – Option 011
Step 2. Measurement Calibration
7
Press
FORWARD:SHORT
.
8
When the analyzer displays the message
CONNECT STD THEN PRESS KEY
TO MEASURE
, connect the 50 ohm termination (supplied in the calibration kit)
to reference test port 1, as shown in
.
9
Press
FORWARD:LOAD
.
10
When the analyzer finishes measuring the standard, connect the “open” end of
the open/short combination to reference test port 2, as shown in
Figure 5-2
.
11
Press
REVERSE:OPEN
.
12
When the analyzer displays the message
CONNECT STD THEN PRESS KEY
TO MEASURE
, connect the “short” end of the open/short combination to
reference port 2 as shown in
.
13
Press
REVERSE:SHORT
.
14
When the analyzer finishes measuring the standard, connect the 50 ohm
termination to reference port 2, as shown in
.
15
Press
REVERSE:LOAD
.
16
When the analyzer finishes measuring the standard, press
STANDARDS DONE
.
17
Connect the test port cables as shown
.
Figure 5-3. Transmission Calibration Setup