85056D
2- 5
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications in
apply to the devices in your calibration kit when
connected with an Agilent precision interface.
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent further certifies that its calibration measurements
are traceable to the United States National Institute of Standards and Technology (NIST)
to the extent allowed by the institute’s calibration facility, and to the calibration facilities
of other International Standards Organization members. See
Devices in Your Kit” on page 4-2
Table 2-3 Electrical Specifications for 85056D 2.4 mm Devices
Device
Specification
Frequency (GHz)
Broadband loads
Return loss
≥
42
d
Β (ρ ≤
0.00794)
dc to
≤
4
(male and female)
Return loss
≥
34 d
Β (ρ ≤
0.01995)
> 4 to
≤
20
Return loss
≥
30 dB
(ρ ≤
0.03162)
> 20 to
≤
26.5
Return loss
≥
26 dB
(ρ ≤
0.05019)
> 26.5 to
≤
50
Adapters
Return loss
≥
32 d
Β (ρ ≤
0.02512)
dc to
≤
4
Return loss
≥
30d
Β (ρ ≤
0.03162)
> 4 to
≤
26.5
Return loss
≥
25dB
(ρ ≤
0.05623)
> 26.5 to
≤
40
Return loss
≥
20 dB
(ρ ≤
0.10000)
> 40 to
≤
50
Offset opens
a
a. The specifications for the opens and shorts are given an allowed deviation from the
nominal model as defined in the standard definitions (see
Standard Definitions Values are Available on the Web” on page A-2
).
±
0.5
°
deviation from nominal
dc to
≤
2
(male and female)
±
1.25
°
deviation from nominal
> 2 to
≤
20
±
1.75
°
deviation from nominal
> 20 to
≤
40
±
2.25
°
deviation from nominal
> 40 to
≤
50
Offset shorts
±
0.50
°
deviation from nominal
dc to
≤
2
(male and female)
±
1.25
°
deviation from nominal
> 2 to
≤
20
±
1.5
°
deviation from nominal
> 20 to
≤
40
±
2.0
°
deviation from nominal
> 40 to
≤
50
Содержание 85056D
Страница 4: ...iv 85056D...
Страница 6: ...vi 85056D Contents...
Страница 7: ...1 1 1 General Information...
Страница 15: ...2 1 2 Specifications...
Страница 20: ...2 6 85056D Specifications Electrical Specifications...
Страница 21: ...3 1 3 Use Maintenance and Care of the Devices...
Страница 30: ...3 10 85056D Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 2 4 mm Connectors...
Страница 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Страница 38: ...4 4 85056D Performance Verification Recertification...
Страница 39: ...5 1 5 Troubleshooting...
Страница 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Страница 45: ...6 1 6 Replaceable Parts...
Страница 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Страница 49: ...A 1 A Standard Definitions...