85056D
4- 3
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
NOTE
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a
Standard
calibration for the recertification of the kit. For
more information, contact Agilent Technologies. See
“Contacting Agilent” on page 5-5
.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage
and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is
first used
after
the recertification date.
Where to Send a Kit for Recertification
Contact Agilent Technologies for information on where to send your kit for recertification.
See
“Contacting Agilent” on page 5-5
. Refer to
“Returning a Kit or Device to Agilent” on
for details on sending your kit.
Содержание 85056D
Страница 4: ...iv 85056D...
Страница 6: ...vi 85056D Contents...
Страница 7: ...1 1 1 General Information...
Страница 15: ...2 1 2 Specifications...
Страница 20: ...2 6 85056D Specifications Electrical Specifications...
Страница 21: ...3 1 3 Use Maintenance and Care of the Devices...
Страница 30: ...3 10 85056D Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 2 4 mm Connectors...
Страница 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Страница 38: ...4 4 85056D Performance Verification Recertification...
Страница 39: ...5 1 5 Troubleshooting...
Страница 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Страница 45: ...6 1 6 Replaceable Parts...
Страница 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Страница 49: ...A 1 A Standard Definitions...