3-12
85056A
Use, Maintenance, and Care of the Devices
Gaging Connectors
e. Release the center conductor pullback mechanism and move the body of the sliding
load toward the gage to mate the outer conductor of the sliding load connector with
the outer conductor of the gage connector.
f. Without turning the gage or the sliding load, connect the gage to the sliding load
being measured by interconnecting the male and female connectors. Connect the nut
finger-tight. Do not overtighten.
g. Using an open-end wrench to keep the device body from rotating, use the torque
wrench included in the kit to tighten the connecting nut to the specified torque. Refer
to
“Final Connection Using a Torque Wrench” on page 3-15
for additional
information.
CAUTION
Always move the center conductor pullback mechanism back before locking
the handle. Do not force the handle past the locked position.
h. Move the center conductor pullback mechanism back (away from the connector end
of the sliding load), and place the pullback handle in its locked position.
i. Gently tap the barrel of the gage with your finger to settle the gage reading.
j. Read the gage indicator dial. Read
only
the black
±
signs;
not
the red
±
signs.
For maximum accuracy, measure the connector a minimum of three times and take
an average of the readings. Use different orientations of the gage within the
connector. After each measurement, rotate the gage a quarter-turn to reduce
measurement variations that result from the gage or the connector face not being
exactly perpendicular to the center axis.
k. Compare the average reading with the observed pin depth limits in
. If the pin depth is outside the limits, it must be adjusted before proceeding.
Refer to
“Adjusting the Sliding Load Pin Depth” on page 3-13
Figure 3-3
Gaging the 2.4 mm Sliding Loads
Содержание 85056A
Страница 4: ...iv ...
Страница 7: ...1 1 1 General Information ...
Страница 13: ...2 1 2 Specifications ...
Страница 19: ...3 1 3 Use Maintenance and Care of the Devices ...
Страница 21: ...85056A 3 3 Use Maintenance and Care of the Devices Electrostatic Discharge Figure 3 1 ESD Protection Setup ...
Страница 28: ...3 10 85056A Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 2 4 mm Connectors ...
Страница 39: ...4 1 4 Performance Verification ...
Страница 42: ...4 4 85056A Performance Verification Recertification ...
Страница 43: ...5 1 5 Troubleshooting ...
Страница 47: ...6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85056A Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85056A Calibration Kit ...
Страница 51: ...85056A 6 5 Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85056A Calibration Kit ...
Страница 52: ...6 6 85056A Replaceable Parts Introduction ...
Страница 53: ...A 1 A Standard Definitions ...
Страница 70: ...A 18 85056A Standard Definitions Nominal Standard Definitions ...