3-18
85056A
Use, Maintenance, and Care of the Devices
Connections
Figure 3-7
Connecting the Sliding Load
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces, always support the
devices and connections.
CAUTION
Do
not
turn the device body. Only turn the connector nut. Damage to the
center conductor can occur if the device body is twisted.
If disconnecting a sliding load, leave the center conductor pullback handle in the locked
position.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connecting nut.
3. Complete the separation by hand, turning only the connecting nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the
connectors.
Содержание 85056A
Страница 4: ...iv ...
Страница 7: ...1 1 1 General Information ...
Страница 13: ...2 1 2 Specifications ...
Страница 19: ...3 1 3 Use Maintenance and Care of the Devices ...
Страница 21: ...85056A 3 3 Use Maintenance and Care of the Devices Electrostatic Discharge Figure 3 1 ESD Protection Setup ...
Страница 28: ...3 10 85056A Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 2 4 mm Connectors ...
Страница 39: ...4 1 4 Performance Verification ...
Страница 42: ...4 4 85056A Performance Verification Recertification ...
Страница 43: ...5 1 5 Troubleshooting ...
Страница 47: ...6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85056A Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85056A Calibration Kit ...
Страница 51: ...85056A 6 5 Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85056A Calibration Kit ...
Страница 52: ...6 6 85056A Replaceable Parts Introduction ...
Страница 53: ...A 1 A Standard Definitions ...
Страница 70: ...A 18 85056A Standard Definitions Nominal Standard Definitions ...