![Agilent Technologies 85024A Скачать руководство пользователя страница 57](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925057.webp)
Chapter 7
7-11
Service
Replacement Procedure
Replacing the Type-N Output Connector
Tools Required:
9/16 inch open-end wrench
1. Remove the connector body with the open-end wrench.
2. Remove the male center conductor assembly from the gold plated inner connector. Use your fingers
to pull out the assembly. The small insulator may or may not come out attached to the center
conductor assembly. If it does, remove it from the old center conductor assembly and place it on the
new one. If it is not attached to the center conductor assembly (it remained inside the inner
connector), proceed to step 3.
3. Hold the new center conductor assembly by the large round insulator, not by the gold plated center
conductor. Carefully align the new center conductor and insert it into the inner connector. Press
gently until you feel it click into place.
4. Place the new connector body and tighten with the open-end wrench.
Figure 7-9
Output Connector Parts
NOTE
“Replaceable Parts” on page 6-4
for part numbers.
Содержание 85024A
Страница 4: ...iv ...
Страница 7: ...1 1 1 General Information ...
Страница 13: ...2 1 2 Accessories ...
Страница 19: ...3 1 3 Installation ...
Страница 24: ...3 6 Chapter3 Installation Returning the Product for Service ...
Страница 25: ...4 1 4 Operation ...
Страница 30: ...4 6 Chapter4 Operation Operator s Check ...
Страница 31: ...5 1 5 Performance Tests ...
Страница 40: ...5 10 Chapter5 Performance Tests Average Noise Level ...
Страница 41: ...6 1 6 Replaceable Parts ...
Страница 46: ...6 6 Chapter6 Replaceable Parts Parts Lists ...
Страница 47: ...7 1 7 Service ...
Страница 60: ...7 14 Chapter7 Service Replacement Procedure Figure 7 11 Regulator Parts and Wiring ...
Страница 64: ...7 18 Chapter7 Service Connector Inspection and Cleaning ...