2-19
Channel Setup Menu
Calibrate
Cal status
The Cal Status function displays a screen similar to
.
Key Path
Channel,
Calibrate,
Cal Status
Figure 2-3. A typical Cal Status display.
Current Date
This is the current date and time. You can compare this to the last plug-in
module calibration time. That way you will know how long it has been since
the last plug-in module calibration was performed.
Current Frame
∆
Temp
This is the temperature change on the inside of the instrument since the last
mainframe calibration was performed. A positive number indicates how many
degrees warmer the mainframe is currently as compared to the temperature
of the mainframe at the last mainframe calibration.
Channel 1
Calibration
Status
The instrument displays
Calibrated
or
Uncalibrated
, depending on
whether the last plug-in module calibration is still valid. A calibration can be
invalidated if:
• The mainframe has cycled power.
• The plug-in has been repaired, reprogrammed, or removed from the
mainframe.
• The instrument’s operating temperature has changed and remains more than
5
°
C from the temperature at which the Plug-in calibration was performed.
Uncalibrated
indicates the plug-in module vertical calibration is invalid.
Содержание 83482A
Страница 1: ...Agilent 83482A Optical Electrical Plug In Module User s Guide ...
Страница 6: ......
Страница 8: ......
Страница 16: ......
Страница 27: ...2 11 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Страница 28: ...2 12 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Страница 65: ...4 9 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 66: ......
Страница 84: ......
Страница 88: ......