background image

2-11

Using the Agilent 83433A

Connecting the Agilent 83433A to an Oscilloscope

Connecting the Agilent 83433A to an 
Oscilloscope

The following procedure describes how to connect the lightwave transmitter 
to an oscilloscope. The following equipment is used:

• Agilent 83433A lightwave transmitter
• Pattern generator
• Clock source
• Optical multimeter
• Optical oscilloscope

1

Follow the procedure described i

“Performing a Quick Confidence Check” on 

page 2-6

2

After completing this check, connect the 

OPTICAL OUTPUT

 of the Agilent 83433A 

to the optical input of the oscilloscope.

3

Connect the trigger output of the pattern generator to the trigger input of the 
oscilloscope.

4

Evaluate the 

OPTICAL OUTPUT 

of the Agilent 83433A with the OC-192/STM-64 

filters and eye mask selected.

Note

Before starting, be sure to clean all connectors and optical interfaces using the proce-
dures describe in 

“Cleaning Connections for Accurate Measurements” on page 3-6

.

Содержание 83433A

Страница 1: ...underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos In...

Страница 2: ...Agilent 83433A Lightwave Transmitter User s Guide...

Страница 3: ...o defects resulting from improper or inadequate maintenance by Buyer Buyer supplied software or interfac ing unauthorized modifica tion or misuse operation outside of the environmental specifications...

Страница 4: ...al case parts Do not attempt to clean internally WAR N I N G This is a Safety Class 1 product provided with a protective earthing ground incorporated in the power cord The mains plug shall only be ins...

Страница 5: ...ilure to ensure adequate earth grounding by not using this cord may cause instrument damage C AUT IO N Do not connect ac power until you have verified the line voltage is correct as described in GENER...

Страница 6: ...825 1 1999 11 and an FDA LASER class 3B according to 21CFR 1040 10 The total power of light energy radiated out of the LASER OUT connector is 8 5 to 11 dBm Opera tor precautions are necessary to main...

Страница 7: ...s sion rates up through 10 7 Gb s The Agilent 83433A is intended for BER test ing with error performance analyzers such as the Agilent 71612B The Mach Zehnder modulator is adjusted for a fixed zero ch...

Страница 8: ...a Bit Error Ratio Test Set 2 9 Connecting the Agilent 83433A to an Oscilloscope 2 11 3 Reference Options 3 2 Replacement Parts 3 3 Front Panel Fiber Optic Adapters 3 4 Power Cords 3 5 Cleaning Connec...

Страница 9: ......

Страница 10: ...ct the Shipment 1 3 Step 2 Check the Fuse 1 5 Step 3 Connect the Line Power Cable 1 6 Step 4 Turn on the Agilent 83433A 1 8 Step 5 Avoid costly repairs 1 9 Step 6 Learn more about our products 1 10 Ge...

Страница 11: ...433A Refer to Chapter 3 Reference for the following addi tional information Tips on avoiding costly repairs by proper optical connection cleaning techniques Lists of available accessories and power co...

Страница 12: ...he shipment for completeness and have checked the light wave transmitter mechanically and electrically The lightwave transmitter is packed within a carton Refer to Returning the Instrument for Service...

Страница 13: ...kes frequent improvements to its products to enhance their performance usability or reliability and to control costs Agi lent service personnel have access to complete records of design changes to eac...

Страница 14: ...open the fuse holder door 4 The fuse is housed in a small container Insert the tip of a screwdriver on the side of the container and gently pull outward to remove the container A spare fuse is stored...

Страница 15: ...ument damage C AUT IO N Do not connect ac power until you have verified the line voltage is correct as described in the following paragraphs Damage to the equipment could result C AUT IO N This instru...

Страница 16: ...ct the Agilent 83433A to ac power outlets unique to specific geographic areas The cable appropriate for the area to which the Agilent 83433A is originally shipped is included with the unit You can ord...

Страница 17: ...disconnects the mains circuit from the mains supply after the EMC filters and before other parts of the instrument If the Agilent 83433A fails to turn on properly consider the following possibili tie...

Страница 18: ...nstrument repairs damaged cables and compromised measurements Before you connect any fiber optic cable to the Agilent 83433A refer to Cleaning Connections for Accurate Measurements on page 3 6 C AUT I...

Страница 19: ...rn more about Fiber Optic Test Equipment go to the Agilent Technolo gies home page listed above and follow this path 1 Click Products 2 Click Test and Measurement 3 Under Products click Fiber Optic Te...

Страница 20: ...ar Panel Features 2 4 Using a Laser Source 2 5 Performing a Quick Confidence Check 2 6 Connecting the Agilent 83433A to a Bit Error Ratio Test Set 2 9 Connecting the Agilent 83433A to an Oscilloscope...

Страница 21: ...osition before making or removing any optical connections DATA INPUT Modulation input for digital signals The input is AC coupled DATA PRESENT INDICATOR Lights when LASER ENABLE is ON and valid data i...

Страница 22: ...th different adapters as needed refer to Front Panel Fiber Optic Adapters on page 3 4 CLASS IIIB INTERNAL LASER OUT Provides the internal laser output for the instru ment Use only Panda PMF fiber with...

Страница 23: ...2 4 Using the Agilent 83433A Rear Panel Features Rear Panel Features...

Страница 24: ...e rear panel This connector must be shorted for the laser to operate When the short is removed from the con nector the accessible radiation does not exceed the AEL for FDA Laser Class 1 IEC Laser Clas...

Страница 25: ...e equipment as shown in Figure 2 1 Figure 2 1 Setup to perform a quick confidence check 2 Turn on the Agilent 83433A and let it warm up for 30 minutes 3 Turn on the pattern generator and clock source...

Страница 26: ...to the EXTENAL LASER INPUT 7 Connect the 83433A CLASS I OPTICAL OUT to the input of an optical multimeter 8 Set the pattern generator as follows Clock output frequency 9953 28 MHz Data output External...

Страница 27: ...perating correctly 6 Review the procedure for the test being performed when the problem appeared Are all the settings correct 7 Are the connectors clean Refer to Cleaning Connectors on page 3 12 for m...

Страница 28: ...The following procedure describes how to connect the lightwave transmitter to a bit error ratio test set BERT The following equipment is used Agilent 83433A lightwave transmitter Lightwave receiver or...

Страница 29: ...ecting the Agilent 83433A to a bit error ratio test system 2 Use the optical cable with Panda polarization maintaining fiber to connect the Agilent 83433A CLASS IIIb INTERNAL LASER output to the EXTER...

Страница 30: ...low the procedure described in Performing a Quick Confidence Check on page 2 6 2 After completing this check connect the OPTICAL OUTPUT of the Agilent 83433A to the optical input of the oscilloscope 3...

Страница 31: ......

Страница 32: ...lacement Parts 3 3 Front Panel Fiber Optic Adapters 3 4 Power Cords 3 5 Cleaning Connections for Accurate Measurements 3 6 Returning the Instrument for Service 3 17 Agilent Technologies Service Office...

Страница 33: ...Receiver User s Guide Agilent part number 83434 90001 Available seperately The Fiber Optics Handbook Agilent part number 5952 9654 is an introduc tion and reference for fiber optic measurements Optio...

Страница 34: ...Parts Replacement Parts Table 3 2 Replaceable Parts Description Agilent Part Nmmber APC 3 5 F to F connector saver 5061 5311 APC 3 5 50 ohm termination 1810 0118 Optical jumper cable 5022 1843 FC PC c...

Страница 35: ...ent Part Number Diamond HMS 10 81000AI FC PCa a The FC PC adapter is the standard adapter supplied with the instrument 81000FI D4 81000GI SC 81000KI DIN 81000SI ST 81000VI Biconic 81000WI Dust Covers...

Страница 36: ...nations 125V 8120 1378 8120 1521 8120 1992 Straight NEMA5 15P 90 Straight Medical UL544 90 228 90 228 96 244 Jade Gray Jade Gray Black United States Canada Mexico Philippines Taiwan 250V 8120 2104 812...

Страница 37: ...urve and concentricity of the core within the cladding Mating one style of cable to another requires an adapter Agilent Technologies offers adapters for most instruments to allow testing with many dif...

Страница 38: ...s However many instrument specifications require tighter tolerances than most connectors including the FC PC style can deliver These instruments cannot tolerate connectors with the large non concentri...

Страница 39: ...the desired position within 0 2 m This process plus the keyed axis allows very precise core to core alignments This connector is found on most Agilent Technologies lightwave instruments The soft core...

Страница 40: ...ber s 125 m core and cladding which carries the light The surrounding area is the soft nickel silver ferrule Figure 3 5 shows a dirty fiber end from neglect or perhaps improper cleaning Material is sm...

Страница 41: ...5 Dirty fiber end and ferrule from poor cleaning Figure 3 6 Damage from improper cleaning While these often work well on first insertion they are great dirt magnets The oil or gel grabs and holds gri...

Страница 42: ...rs that have a convex fiber end the end can be pushed off axis resulting in misalignment and excessive return loss Many measurements are actually improved by backing off the connector pressure Also if...

Страница 43: ...mmon Visual inspection of fiber ends Visual inspection of fiber ends can be helpful Contamination or imperfections on the cable end face can be detected as well as cracks or chips in the fiber itself...

Страница 44: ...hat index matching compounds not be applied to their instruments and accessories Some compounds such as gels may be difficult to remove and can contain damaging particulates If you think the use of su...

Страница 45: ...es and other parts of the connector while avoiding the end of the fiber 3 Apply isopropyl alcohol to a new clean lint free cotton swab or lens paper 4 Clean the fiber end with the swab or lens paper T...

Страница 46: ...after the initial cleaning seems poor try cleaning the con nector again Often a second cleaning will restore proper performance The second cleaning should be more arduous with a scrubbing action To c...

Страница 47: ...an dry foam swab 4 Blow through the adapter using filtered dry compressed air Nitrogen gas or compressed dust remover can also be used Do not shake tip or invert compressed air canisters because this...

Страница 48: ...0801 If the instrument is still under warranty or is covered by an Agilent Technolo gies maintenance contract it will be repaired under the terms of the warranty or contract the warranty is at the fro...

Страница 49: ...of the instrument Place protective wrap around the cable and tape it to the top panel of the instrument 3 Cover all front or rear panel connectors that were originally covered when you first received...

Страница 50: ...o protect the instrument and prevent it from moving in the carton If packing foam is not available the best alternative is S D 240 Air Cap from Sealed Air Corporation Commerce California 90001 Air Cap...

Страница 51: ...nologies Service Numbers Austria 01 25125 7171 Belgium 32 2 778 37 71 Brazil 11 7297 8600 China 86 10 6261 3819 Denmark 45 99 12 88 Finland 358 10 855 2360 France 01 69 82 66 66 Germany 0180 524 6330...

Страница 52: ...4 Agilent 83433A Specifications and Characteristics 4 3 Regulatory Information 4 6 Specifications and Regulatory Information...

Страница 53: ...ng unless otherwise noted All specifications apply after the instrument s temperature has been stabilized after 30 minutes of continuous operation Specifications pertaining to the use of the internal...

Страница 54: ...gital data input Low frequency modulation input External optical modulator inputb 2 0 V pk pk 2 0 2 0 V dc 2 0 V pk pk 14 dBm Low Frequency Modulation Input Frequency range 3 dB points Amplitude for 5...

Страница 55: ...characteristic 3 5 kg 7 8 lb Dimensions 102mm 4 in height 216 mm 8 5 in width 444 mm 17 5 in depth Agilent System II half width case FRONT PANEL INPUT OUTPUT Digital data input connector APC 3 5 male...

Страница 56: ...n 10 dBm may cause modulated optical output power to exceed Class I level c Class IIIB laser output d Wavelength adjustment disabled e Average power with internal laser modulated with 231 1 PRBS f Mea...

Страница 57: ...designed for use in INSTALLATION CATEGORY II and POLLU TION DEGREE 2 per IEC 61010 1 and 664 respectively Notice for Germany Noise Declaration This is to declare that this instrument is in conformance...

Страница 58: ...4 7 Specifications and Regulatory Information Regulatory Information Declaration of Conformity...

Страница 59: ......

Страница 60: ...care 3 6 connector interface front panel 3 2 cotton swabs 3 14 D DATA 2 2 data input connector 2 2 declaration of conformity 4 7 dust caps 3 14 E external laser input 2 3 F fiber optics adapters 2 3...

Страница 61: ...hutdown connector 2 5 repair options 2 8 returning for service 3 17 S safety iii v laser classification iii line fuse iii 1 5 sales and service offices 3 20 SDH SONET compliance vi serial numbers 1 4...

Страница 62: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...