Maintenance
5
SCD and NCD User Manual
75
10
Reinstall the transfer line onto the outlet fitting. Tighten
until snug (finger-tight). Do not overtighten.
11
Restore the detector gas flows.
12
Check for leaks at the upper hydrogen fitting. Correct a leak
as needed.
13
Restore the remaining detector operating conditions.
14
Reset the EMF counter.