Performance Test Instructions
Performance Tests
Agilent 81662A and 81663A DFB Laser Source Modules User’s Guide, Fifth Edition
37
All Wavelength Tests
Make sure that the cable connector, detectors and adapters are clean.
Connect the DFB Laser Source Module to the Wavelength Meter as
shown in Figure 8.
Figure 8
Test Setup for Wavelength Tests
General Settings for
Burleigh WA-1500 Wavelength Meter
Set up the Burleigh WA-1500 Wavelength Meter as follows:
• Set
Display
to
Wavelength
• Set
Medium
to
Vacuum
• Set
Resolution
to
Auto
• Set
Averaging
to
On
• Set
Input Attenuator
to
Auto
.
Wavelength Repeatability Test
1
Move to the DFB-Source channel of the Agilent 8163A/4A/6A
mainframe and press
Menu
.
2
Set the
Wavelength mode
to
Offset
and adjust the
Power
reading to
1mW or 0dBm
3
Close
the Menu.
4
Press the key beside the laser output to switch on the laser ouput.
5
Adjust the frequency offset to the first value given in the test record
in column "from wavelength".
Burleigh Wavemeter WA-1500
WA 1500
Burleigh
burleigh
Mainframe 8163A w/ 8166xA Source
8163A
8166xA
81113PC
Содержание 81662A
Страница 2: ...S1 Agilent 81662A and Agilent 81663A DFB Laser Source Modules User s Guide ...
Страница 13: ...Table of Contents 12 Agilent 81662A and 81663A DFB Laser Source Modules User s Guide Fifth Edition ...
Страница 20: ...Agilent 81662A and 81663A DFB Laser Source Modules User s Guide Fifth Edition 19 Accessories ...
Страница 26: ...Agilent 81662A and 81663A DFB Laser Source Modules User s Guide Fifth Edition 25 Specifications ...
Страница 34: ...Agilent 81662A and 81663A DFB Laser Source Modules User s Guide Fifth Edition 33 Performance Tests ...
Страница 69: ...68 Agilent 81662A and 81663A DFB Laser Source Modules User s Guide Fifth Edition ...
Страница 70: ......