Performance Tests
Performance Tests
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
55
Return Loss Test
This performance test applies to Agilent 81622B/3B/4B/6B/7B Optical
Heads.
1
Make sure that all connectors are clean.
2
Connect the equipment as shown in Figure 13.
3
Press
[Preset]
on the mainframe.
Figure 13
Return Loss Reference Setup
N O T E
To ensure traceability, use the 81610CC Reference Cable for
calibration measurements. Do
not
use the 81610CC Reference Cable
for measurements on a Device Under Test. Instead, use a measurement
patchcord.
C A U T I O N
It is important to maintain the quality of the straight connector end of
the 81610CC Reference Cable. Never add another connector to the
straight end of the 81610CC Reference Cable, since a physical
connection is made. When the straight end of the 81610CC Reference
Cable is connected to the 81634B Power Sensor module, no physical
connection is made so there should be no degradation of connector
quality.
4
Zero the Return Loss Module and the Optical Head
5
Enable the internal laser source of the Return Loss Module and
allow 20 minutes for it to stabilize.
6
At the Power Meter:
a
Set the averaging time
[AvTime]
to 1s
Содержание 81618A
Страница 9: ...Table of Contents 8 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 16: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 15 Accessories ...
Страница 22: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 21 Specifications ...
Страница 36: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 35 Performance Tests ...
Страница 88: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 87 Cleaning Information ...
Страница 111: ...110 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 112: ......
Страница 113: ... Agilent Technologies GmbH 2001 2002 Printed in Germany February 20 2002 81623 90B12 81623 90B12 www agilent com ...