Pulse Trailing-Delay Parameter
The trailing delay is the absolute delay from the start of the pulse period
to the start of the trailing edge. In this format the trailing edge remains
fixed relative to the start of the pulse period if you adjust the pulse-delay
(leading-edge delay) or the pulse period.
Pulse Leading-Edge Parameter
Set the transition time for the leading edge of the pulse, measured from
10% to 90% of the pulse amplitude.
There are two formats available (selectable in text mode):
L
EAD
E
D
The absolute transition time measured from 10% to 90% of pulse
amplitude. In this format the leading edge is independent of the pulse
width.