General Maintenance
To Install a Filament on the EI XTR Source
7000/7010 Series TQ GC/MS Operating Manual
191
To Install a Filament on the EI XTR Source
Materials needed
•
Filament assembly, EI (G7005-60061)
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
Procedure
1
Remove the old filament. (See
“To Remove a Filament from the EI XTR
2
Place the new filament into its position on the ion source body. (See
3
Secure the filament to the ion source body with the screw.
4
After installing the filament, verify that it is not grounded to source body.
5
Reinstall the ion source. (See
“To Install the EI XTR Source”
6
Close the front analyzer chamber. (See
7
Pumpdown the MS. (See
8