6
Operating in CI Mode
To Pumpdown and Switch from EI to CI Mode
146
7000/7010 Series TQ GC/MS Operating Manual
14
Click
Autotune
to start the autotune. At the completion of the autotune a
Tune Report is printed.
15
Review the Tune Report. If the results are acceptable, save the autotune by
clicking the
Files and Report
tab, then click
Save
. Default tune control limits
are shown in
Notes for
•
N/A: Not available.
•
Abundance target: Adjust higher or lower to get desired signal abundance.
Higher signal abundance also gives higher noise abundance. This is adjusted
for data acquisition by setting the EMV in the method.
•
Peakwidth target: Higher peakwidth values give better sensitivity, lower
values give better resolution.
•
Maximum emission current: Optimum emission current maximum for NCI is
very compound-specific and must be selected empirically. Optimum
emission current for pesticides, for example, may be about 200 µA.
Table 13 Default Tune Control Limits, used by CI autotune only
Reagent gas
Methane
Ammonia
Ion polarity
Positive
Negative
Positive
Negative
Abundance target
1×10
6
1×10
6
N/A
1×10
6
Peak width target
0.7
0.7
N/A
0.7
Maximum repeller
4
4
N/A
4
Maximum emission current, µA
240
50
N/A
50
Max electron energy, eV
240
240
N/A
240
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...