To Set the GC/MS Interface Temperature from MassHunter
7000/7010 Series TQ GC/MS Operating Manual
111
To Set the GC/MS Interface Temperature
from MassHunter
You can also use the
GC Control
panel to perform this task.
Procedure
1
Select
Instrument > GC Parameters from the Instrument Control panel.
2
Click the
Aux Heaters
icon to edit the interface temperature. (See
example has the GC/MS interface temperature configured as Thermal Aux 2.
3
Click the heater
On
check box, and type the setpoint in the
Value °C
column.
The typical setpoint is 280 °C. The limits are 0 °C to 400 °C. A setpoint below
ambient temperature turns off the interface heater.
4
Click
Apply
to download setpoints or click
OK
to download setpoints, and
close the window.
5
To make the new settings part of the method, select
Save
from the
Method
menu.
Figure 33.
Setting the interface temperature
CAUTION
Ensure that the carrier gas is turned on and the column has been purged of air
before heating the GC/MS interface or the GC oven.
CAUTION
When setting the GC/MS interface temperature, never exceed the maximum
for your column.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...