
5
Operating in EI Mode
To Set the MS Analyzer’s Temperatures
7000/7010 Series TQ GC/MS Operating Manual
109
To Set the MS Analyzer’s Temperatures
Setpoints for the MS ion source, front quad (MS1), rear quad (MS2), and
temperatures are stored in the current tune file. When a method is loaded, the
setpoints in the tune file associated with that method are downloaded
automatically.
Procedure
1
In
Instrument Control
view, select the
MS Tune
icon to display the
Tune
dialog box. Select the
Manual Tune
tab, then select the
Ion Source
tab to
display the ion source parameters.
2
Enter the temperature setpoint in the
Source Temp
for
recommended setpoints.
3
Select the
MS1
tab to display the MS1 parameters. (See
4
Enter the temperature setpoint in the
MS1 Quad Temp
field. See
for
the recommended setpoint.
5
Select the
MS2
tab to display the MS2 parameters.
6
Enter the temperature setpoint in the
MS2 Quad Temp
field. See
for
recommended setpoint.
7
Select the
Files and Reports
tab, then click
Save
to save the tune file with
these changes.
Table 8
Recommended temperature settings
Zone
EI operation
MS Source
230 °C default. High limit 350 °C. Application dependent.
MS Quad 1
150 °C
MS Quad 2
150 °C
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...