146
Troubleshooting and Maintenance Manual
7
Analyzers and Collision Cell
Overview
The complete analyzer system is shown in
The front analyzer
The front analyzer consists of the ion source, the front quadrupole,
heaters/sensors, and the post-filter.
The sample components exiting the GC column flow into the ion source. In the
ion source, the sample molecules are ionized and fragmented. The resulting
ions are repelled from the ion source into the front quadrupole mass filter. The
selected ions pass through the filter, and post-filter, and enter the collision
cell.
The post-filter
This device, located on the end of the front quadrupole, helps transfer the
selected ions to the collision cell. It also aligns the front of the collision cell
with the front analyzer.
Figure 26
The analyzer system
Front analyzer (MS1)
Post-filter Collision cell
Pre-filter
Rear analyzer (MS2) Detector
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...