Agilent 6890 Gas Chromatograph Service Manual
Gases for capillary columns
When used with capillary columns, GC detectors require a separate makeup
gas for optimum sensitivity. For each detector and carrier gas, there is a
preferred choice for makeup gas. Table 520-2 lists gas recommendations for
capillary columns.
Table 520-2
Gas Recommendations for Capillary Columns
Detector
Carrier
gas
Preferred
makeup gas
Second choice
Detector anode
purge or reference
gas
Electron Capture
Hydrogen
Argon/Methane
Nitrogen
Anode purge must
be same as makeup
Helium
Argon/Methane
Nitrogen
Nitrogen
Nitrogen
Argon/Methane
Argon/
Methane
Argon/Methane
Nitrogen
Flame Ionization
Hydrogen
Nitrogen
Helium
Hydrogen and
air for detector
Helium
Nitrogen
Helium
Nitrogen
Nitrogen
Helium
Flame
Photometric
Hydrogen
Nitrogen
Hydrogen and air for
detector
Helium
Nitrogen
Nitrogen
Nitrogen
Argon
Nitrogen
Nitrogen-
Phosphorus
Helium
Nitrogen
Helium**
Hydrogen and
air for detector
Nitrogen
Nitrogen
Helium**
Thermal
Conductivity
Hydrogen*
Must be same
as carrier and
reference gas
Must be same
as carrier and
reference gas
Reference must be
same as carrier and
makeup
Helium
Nitrogen
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......