1 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
520
Gases
Gas requirements
Gases for packed columns
The carrier gas you use depends upon the type of detector and the
performance requirements. Table 520-1 lists gas recommendations for
packed column use. In general, makeup gases are not required with packed
columns.
Table 520-1
Gas Recommendations for Packed Columns
Detector
Carrier Gas
Comments
Detector anode purge
or reference gas
Electron Capture
Nitrogen
Maximum sensitivity
Nitrogen
Argon/
Methane
Maximum dynamic range
Argon/Methane
Flame Ionization
Nitrogen
Maximum sensitivity
Hydrogen and
air for detector
Helium
Acceptable alternative
Flame
Photometric
Hydrogen
Hydrogen and air for
detector
Helium
Nitrogen
Argon
Nitrogen-
Phosphorus
Helium
Optimum performance
Hydrogen and
air for detector
Nitrogen
Acceptable alternative
Thermal
Conductivity
Helium
General use
Reference must be same as
carrier
Hydrogen
Maximum sensitivity (Note A)
Nitrogen
Hydrogen detection (Note B)
Argon
Maximum hydrogen sensitivity
(Note B)
Note A: Slightly greater sensitivity than helium. Incompatible with some compounds.
Note B: For analysis of hydrogen or helium. Greatly reduces sensitivity for other compounds.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......