Operation Verification 109
Table B-4. Operation Verification Test Parameters for Series 664xA
Test Description
Minimum Spec
Results *
Maximum
Spec
Measurement
Uncertainty
MODEL Agilent 6641A
Voltage Programming and Readback
Low Voltage (0 V) V
out
-5 mV
____________ mV
+ 5 mV
2.0
µ
V
Front Panel Display Readback
Vout -6.0 mV
____________ mV
Vout +6.0 mV
2.0
µ
V
High Voltage (8 V) V
out
7.990 V
_____________V
8.010 V
88
µ
V
Front Panel Display Readback
Vout -11.6 mV
____________ mV
Vout +11.6 mV
88
µ
V
Current Programming and Readback
Low Current (0 A) I
out
-26 mA
____________ mA
+26 mA
153
µ
A
Front Panel Display Readback
Iout -18 mA
____________ mA
Iout +18 mA
153
µ
A
High Current (20 A) I
out
19.944 A
____________ A
+20.056 A
2.7 mA
Front Panel Display Readback
Iout -48 mA
____________ mA
Iout +48 mA
2.7 mA
MODEL Agilent 6642A
Voltage Programming and Readback
Low Voltage (0 V) V
out
-10 mV
____________ mV
+ 10 mV
2.0
µ
V
Front Panel Display Readback
Vout -15 mV
____________ mV
Vout +15 mV
2.0
µ
V
High Voltage (20 V) V
out
19.978 V
_____________V
20.022 V
335
µ
V
Front Panel Display Readback
Vout -29 mV
____________ mV
Vout +29 mV
335
µ
V
Current Programming and Readback
Low Current (0 A) I
out
-13 mA
____________ mA
+13 mA
20
µ
A
Front Panel Display Readback
Iout -9.1 mA
____________ mA
Iout +9.1 mA
20
µ
A
High Current (10 A) I
out
9.972 A
____________ A
+10.028 A
3.1 mA
Front Panel Display Readback
Iout -24.1 mA
____________ mA
Iout +24.1 mA
3.1 mA
MODEL Agilent 6643A
Voltage Programming and Readback
Low Voltage (0 V) V
out
-15 mV
____________ mV
+ 15 mV
2.0
µ
V
Front Panel Display Readback
Vout -25 mV
____________ mV
Vout +25 mV
2.0
µ
V
High Voltage ( 35 V) V
out
34.964 V
_____________V
35.036 V
525
µ
V
Front Panel Display Readback
Vout -50 mV
____________ mV
Vout +50 mV
525
µ
V
Current Programming and Readback
Low Current (0 A) I
out
-6.7 mA
____________ mA
+6.7 mA
16
µ
A
Front Panel Display Readback
Iout -5 mA
____________ mA
Iout +5 mA
16
µ
A
High Current ( 6 A) I
out
5.985 A
____________ A
6.015 A
1.1 mA
Front Panel Display Readback
Iout -14 mA
____________ mA
Iout +14 mA
1.1 mA
MODEL Agilent 6644A
Voltage Programming and Readback
Low Voltage (0 V) V
out
-26 mV
____________ mV
+ 26 mV
2.0
µ
V
Front Panel Display Readback
Vout -40 mV
____________ mV
Vout +40 mV
2.0
µ
V
High Voltage ( 60 V) V
out
59.938 V
_____________V
60.062 V
845
µ
V
Front Panel Display Readback
Vout -82 mV
____________ mV
Vout +82 mV
845
µ
V
Current Programming and Readback
Low Current (0 A) I
out
-4.1 mA
____________ mA
+4.1 mA
16
µ
A
Front Panel Display Readback
Iout -3 mA
____________ mA
Iout +3 mA
16
µ
A
High Current (3.5 A) I
out
3.491 A
____________ A
+3.509 A
500
µ
A
Front Panel Display Readback
Iout -8.3 mA
____________ mA
Iout +8.3 mA
500
µ
A
* Enter your test results in this column.
Содержание 6641A
Страница 50: ......
Страница 94: ......
Страница 97: ...Calibration 97 Figure A 1 Calibration Test Setup c Series 668xA 669xA Setup ...
Страница 106: ...Operation Verification 106 Figure B 1 Verification Test Setup c Series 668xA 669xA Setup ...
Страница 118: ......
Страница 128: ......
Страница 131: ...Current Loop Compensation Series 668xA Only 131 Figure E 1 CC Loop Compensation Curves For Models 6682A and 6683A ...
Страница 136: ......