56
Table 3-10. Performance Failure Symptoms
DEFECTIVE
SYMPTOMS
BOARD
CHECK FUNCTIONAL CIRCUITS
unexplained OVP shutdowns
A2
OVP circuit, CV circuit
no current limit
A2
CC circuit
max current < specified
A2
CC Clamp, CC circuit
max power < specified
A2, A1
Power Limit, 20KHz clock, transformer AlT1
max voltage < specified
A2, A1
CV Circuit, diodes A1CR1-CR4
cycles on & off randomly
A2, A1
AC-Surge-&-Dropout Detector, Mains Voltage
Select switch A1S2
CV overshoots
A2
A2U5A, A2CR19, A2R62
output noise (<1KHz)
A2,A1
CV circuit, input filter
output noise (>1KHz)
A1, A4
transformer AlT2, Output Filter, snubbers A4R1 to
A4R11, A4R13 to A4R19, A4C1 to A4C4, A4CR2,
A4CR3, A2R15, A2C2
CV regulation, transient
A2, A1
wrong sensing
response, programming time
low ac mains voltage, CV circuit
CC regulation
A2
low ac mains voltage, CC circuit
CV oscillates with capacitive
A2
A2R61, A2R60, A2R58, A2R59, A2C33, A2R64,
loads
A2R68, A2C36, A2C37, A2U5, A2R65
CC oscillates with inductive
A2
A2R61, A2R60, A2R58, A2R57, A2C33, A2R19
loads
A2C11, A2R28, A2C12, A2U4, A2R35, A2C20,
A2U4, A2R37, A2C17, A2R29, A2C18, A2R31
Table 3-11. No-Output Failures
(Bias supplies and AC turn-on circuit functioning)
Status of FET On/Off-Pulses
PWM-ON
A2J7-26
PWM-OFF
A2J7-25
DEFECTIVE
BOARD
CHECK FUNCTIONAL CIRCUITS
Lo
Lo
A2
Control ckts: CV & CC thru On- & Off-Pulse Oneshots *
Lo
Hi
A2&A4
PWM and DC-to-DC Converter: A4 PFETS probably failed
Hi
Lo
A2&A4
PWM and DC-to-DC Converter: A4 PFETS probably failed
Hi
Hi
A2&A4
PWM and DC-to-DC Converter: A4 PFETS probably failed
Lo
N
A2
A2U15A, On-Pulse Oneshot and Q11
N
Lo
A2&A4
Off-Pulse Oneshot and DC-to-DC: A4 PFETS probably failed
Hi
N
A2&A4
A2U15A, On-Pulse Oneshot & DC-to-DC: A4 PFETS probably failed
N
Hi
A2&A4
Off-Pulse Oneshot and DC-to-DC: A4 PFETS probably failed
N
N
A2&A4
Power-Limit Comparator and DC-to-DC: A4 PFETS probably failed
Lo= TTL low
Hi= TTL high
N= normal 20KHz pulse train, TTL levels
* Decide which to troubleshoot--the CV circuit, the CC circuit, or the PWM and Off-Pulse & On-Pulse Oneshots-- by
measuring the CV CONTROL (A2CR24, cathode) and the CC CONTROL (A2CR11 cathode) voltages. Troubleshoot
whichever is negative, and if neither is negative, troubleshoot the PWM. Make these voltage measurements after you
have implemented the Main Troubleshooting Setup.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 6030A
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 33: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 114: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 115: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 116: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 117: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 125: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...