8
CI Maintenance
To Install the CI Source
250
5977B Series MSD Operating Manual
To Install the CI Source
Procedure
The video shows wiring connections for the non-HES model MSD.
1
Vent the MSD. (See
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Slide the CI source into the radiator. (See
4
Install the thumbscrews. (See
5
Connect the wiring to the CI source. (See
“To Connect/Disconnect non-HES
Model Wiring from the CI Source”
EI HES Model Wiring to the CI Source”
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap and take other antistatic precautions
before
you open the
analyzer chamber.
Figure 82.
Installing the CI source
Ion source
Thumbscrews
Source radiator
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...