8
CI Maintenance
To Install a CI Source Filament
5977B Series MSD Operating Manual
249
To Install a CI Source Filament
Materials needed
•
Filament assembly, 2-pk, CI (G7005-60072)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)Tweezers (8710-2460)
Procedure
1
Remove the old filament. (See
“To Remove the CI Source Filament”
2
Place the new filament into its position on the ion source body. (See
3
Secure the filament to the ion source body with the screw. (See
4
After installing the filament, verify that it is not grounded to source body.
5
Reinstall the CI source. (See
6
Pumpdown the MSD. (See
“To Pumpdown the MSD in CI Mode”
7
Autotune the MSD.