8
CI Maintenance
To Switch from an EI XTR, SS, or Inert Source to the CI Source
5977B Series MSD Operating Manual
233
To Switch from an EI XTR, SS, or Inert Source
to the CI Source
Procedure
1
Vent the MSD. (See
2
Open the analyzer. (See
“To Open the Analyzer Chamber”
3
Remove the EI source. (See
“To Remove the EI XTR, SS, or Inert Source”
4
If removing an XTR source, remove the brown extractor wire from the
ceramic source board and store it with the EI XTR source. (See
5
Install the CI source. (See
6
Install the CI/Extractor interface tip seal if it is not already installed
(p/n G3870-20542). (See
“To Install the GC/MSD Interface Tip Seal”
page 50.)
7
Close the analyzer. (See
“To Close the Analyzer Chamber”
8
Pumpdown the MSD. (See
“To Pumpdown the MSD in CI Mode”
CAUTION
Always verify MSD performance in EI before switching to CI operation.
Always set up the CI MSD in PCI first, even if you are going to run NCI.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap. (See
on page 171.) Take
antistatic precautions
before
you open the analyzer chamber.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...