7
General Maintenance
To Assemble an EI SS or Inert Source
210
5977B Series MSD Operating Manual
CAUTION
Do not overtighten the interface socket. Overtightening could strip the threads.
Figure 67.
Assembling the EI standard or EI Inert source
1
2
8*
15
15
15
15
14
14
14
2*
2*
16
16
16
8*
12*
12*
12*
10
2*
8*
8*
7*
17
18
3
8
8
7
12
13
4
6
5
11
9
Table 22 Parts list for the EI standard or EI Inert source (
)
Item number
Item description
Part number (SSL)
Part number (Inert)
1
Gold plated set screw
G1999-20022
G1999-20022
2
Gold plated screw
G3870-20021
G3870-20021
3
Interface socket
G1099-20136
G1099-20136
4
Source body
G1099-20130
G2589-20043
5
Drawout cylinder
G1072-20008
G1072-20008
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...