1
Introduction
Vacuum gauge
18
5977B Series MSD Operating Manual
•
GC/MSD interface
•
Electronics
•
Analyzer
Vacuum gauge
The MSD may be equipped (or ordered) with an ion vacuum gauge. The Agilent
MassHunter GC/MS Acquisition software can be used to read the pressure (high
vacuum) in the vacuum manifold. Operation of the gauge controller is described
in this manual. The gauge is
required
for CI operation.
Table 3
5977B Series MSD features
Feature
High vacuum pump
Diffusion
Turbo
Optimal He column flow mL/min
1
1 to 2
Maximum recommended
gas flow mL/min
*
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas
use. For other gases the maximum flow will vary.
1.5
4
Maximum gas flow, mL/min
†
† Expect degradation of spectral performance and sensitivity.
2
6.5
Max column id
0.25 mm (30 m)
0.53 mm (30 m)
CI capability
‡
‡ Turbo pump models are field upgradeable to CI.
No
Yes
Inert ion sources available
Yes
Yes
GC compatibility
9000/8890/
8860/7890 Series
9000/8890/
8860/7890 Series
Foreline pumps available
Pfeiffer Duo
Pfeiffer Duo,
Pfeiffer MVP-070-3x,
Agilent IDP-3 24V
DIP
**
capability (3rd party)
** Direct insertion probe.
Yes
Yes
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...