background image

Agilent 5352B  40 GHz CW Microwave

Counter

Data Sheet

Input 1

Frequency Range: 500 MHz to 40 GHz (500 MHz to 46 GHz with Opt. 005)

Sensitivity
 Full Operating Environment
500 MHz to 26.5 GHz: -25 dBm
26.5 GHz to 46 GHz: dBm = 0.741 f(GHz) -44.6
@25° C (typical)
500 MHz to 26.5 GHz: -30 dBm
26.5 GHz to 46 GHz: dBm = 0.741 f(GHz) -49.6

Maximum Input: +7 dBm

Damage Level: +25 dBm

Input 2

Frequency Range
10 Hz to 525 MHz
50 ohm: 10 MHz to 525 MHz
1 Mohm: 10 Hz to 80 MHz

Sensitivity
Full operating environment
50 ohm: 10 MHz to 525 MHz, 25 mV rms (15 mV typical @ 25° C)
1 Mohm: 10 Hz to 80 MHz, 25 mV rms (15 mV typical @ 25° C)

Maximum Input
50 ohm: + 10 dBm
1 Mohm: 1V rms

Damage Level
50 ohm or 1 Mohm dc to 5 kHz: 250 V (dc + ac peak);
>5 kHz: 5.5 V rms (+28 dBm) + 1.25 X 10^6 Vrms/freq

Connector: Replaceable fuse, type BNC (female)

Resolution: Selectable, 1 Hz to 1 MHz