background image

30,000 Molecular Weight Cut-Off (MWCO)
15 mL Spin Concentrator, Part No. 5188-2799

05/2004A

Instructions – 15 mL Concentrators

Equipment required

15 mL Spin Concentrators, pack of 10

Centrifuge
Rotor type

Swing bucket or Fixed angle

Minimum rotor angle

25°

Rotor cavity

To fit 30 x 114 mm conical bottom tubes or
29 x 104 mm round bottom tubes (50mL)

Minimum effective speed

1,000 g

Recommended speed

2,000 g

Maximum speed

3,000 g

Concentrate recovery
Pipette type

Fixed or variable volume

Recommended tip

Thin gel loader type

Technical specifications

Concentrator sample volume capacity
Swing bucket rotor

2-15 mL

Fixed angle rotor

2-8 mL

Dimensions
Total Length

76 mm

Width

25.5 mm

Active Membrane Area

4 cm

2

Hold-up volume of membrane

<20 µL

Dead stop volume

50 µL

pH range

pH 1 – 9

Materials of construction
Body

Polycarbonate

Filtrate vessel

Polypropylene

Concentrator cap

Polycarbonate

Membrane

Polyethersulfone

Concentrating Proteins with 15 mL Concentrators:

1

 

Fill concentrator with sample within the working volume
capacity range noted in the table above, taking care not to
touch the membrane surface, and then close the hinged lid

(ensure that the lid is fully seated).

2

 

Place concentrator into 50 mL centrifuge tube 

(filtrate

collection vessel – not supplied, user must provide)

.

3

 

Insert assembled concentrator into centrifuge (when fixed
angle rotors are used, angle the concentrators so that the
printed windows face upwards/outwards).  Spin at

recommended speeds, taking care not to exceed the maximum
g force indicated in the table above (20-40 minutes is
common).

4

 

Once the desired concentration is achieved, remove assembly
and recover sample from the bottom of the concentrate pocket
with a pipette.

Desalting/Buffer Exchange

See 4 mL Spin Concentrator instructions on the reverse side.

Notes:

·

 

Spin concentrators are single use disposable
items.  Re-use is not recommended.

·

 

For research use only.

·

 

Select the most appropriate membrane cut-off for
your sample.  For maximum recovery, select a
MWCO at least 50% smaller than the molecular

size of the species of interest.

·

 

These concentrators are designed for use with
biological fluids or aqueous solutions only, not

for organic solvents or strong acids or bases.

·

 

Flow rates are affected by MWCO, porosity,
sample size, concentration, viscosity, centrifuge

force, and temperature.  Adjust centrifuge force
and time accordingly for your samples.

·

 

Do not autoclave.  Sterilize with a 70% ethanol

solution.

For more information or technical assistance, 

Agilent Technologies, Inc.

please call toll free: 1-800-227-9770 or visit

2850 Centerville Road

our web site at: www.agilent.com/chem/bioreagents

Wilmington, DE  19808-1610  USA

Отзывы: