Agilent 4155B/4156B VXIplug&play Driver User’s Guide, Edition 4
2-21
Driver Functions
hp4156b_forcePulse
hp4156b_forcePulse
This function specifies the pulse source (PGU) settings and forces the voltage pulse
immediately. To stop the pulse output, use hp4156b_abortMeasure function.
Syntax
ViStatus _VI_FUNC hp4156b_forcePulse(ViSession vi, ViInt32 channel,
ViInt32 count, ViReal64 base, ViReal64 peak, ViReal64 width, ViReal64 period,
ViReal64 delay, ViReal64 rise, ViReal64 fall);
Parameters
vi
Instrument handle returned from hp4156b_init( ).
channel
Channel number of the pulse generator unit.
27 (PGU1) or 28 (PGU2)
count
Pulse count (number of pulses). 1 to 65535, or 0 (free run
mode).
base
Pulse base value. -40.0 to 40.0 V.
peak
Pulse peak value. -40.0 to 40.0 V.
width
Pulse width. 1E-6 to 9.99 seconds.
period
Pulse period. 1E-6 to 10.0 seconds.
delay
Pulse delay time. 0.0 to 10.0 seconds.
rise
Pulse leading time. 0.1E-6 to 10E-3 seconds.
fall
Pulse trailling time. 0.1E-6 to 10E-3 seconds.
Содержание 4155B
Страница 6: ...6 Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4...
Страница 13: ...Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4 1 Installation...
Страница 20: ...1 8 Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4 Installation Installing 4155B 4156B Driver...
Страница 21: ...Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4 2 Driver Functions...
Страница 95: ...Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4 3 Programming Examples Using HP VEE...
Страница 141: ...Agilent 4155B 4156B VXIplug play Driver User s Guide Edition 4 4 Sample Application Programs for HP VEE...