Amplitude resolution: 0.1 dBm
Absolute amplitude accuracy: ±1 dB
(300 kHz, +10 dBm output)
Dynamic accuracy
(add to absolute accuracy): Add 0.02 dB/dB below +10 dBm
Frequency response ±1 dB
(Variation relative to
level at 300 kHz):
Spurious products
Harmonic products: <-30 dB
Non-harmonic products: <-40 dB
Noise:
<-80dB
Relative to the carrier in a 1 Hz bandwidth for offsets greater
than 500 Hz from the carrier)
Output Port Return Loss:
>20 dB
General Specifications
Environmental
Temperature, standard instrument
Operating: 5 to 50°C
Storage (no disk in drive): -20 to 60°C
Temperature, delete disk option
Operating: 0 to 55°
Storage: -40 to 70°C
Humidity, non-condensing,
standard instrument
Operating: 8% to 80% at 30°C
Storage (no disk in drive): 5% to 95%