A. Introduction
Due to a change of manufacturer of the noise diodes, adjustment of bias current is now necessary to ensure that the
noise source performs within its specified ENR range. This document is supplied with all the following replacement
parts.
CAUTION:
Repair or parts replacement of noise sources should NOT be attempted without full
calibration capability and official field calibration software.
This Installation Note applies to the installation of the following replacement parts:
Table 1
Replacement Parts List
Model
Replacement Bulkhead
Replacement Noise Cartridge
346A Std.
00346-60022
00346-60154
346A Opt. 001
00346-60023
00346-60154
346A Opt. 002
00346-60024
00346-60154
346A Opt. 004
00346-60025
00346-60154
346B Std.
00346-60026
00346-60154
346B Opt. 001
00346-60018
00346-60154
346B Opt. 002
00346-60019
00346-60154
346B Opt. 004
00346-60020
00346-60154
N4000A Std.
N4000-60014
00346-60154
N4000A Opt 001
N4000-60015
00346-60154
N4001A Std.
N4001-60003
00346-60154
N4001A Opt 001
N4001-60004
00346-60154