22
Chapter 1
General Information
Operation
Figure 1-9
Typical example of power reading with Noise Source Off
Step 23. If there is not a change in value, return the 346A/B/C to Agilent
Technologies for repair.
Operator’s Maintenance
Proper connector care is a vital part of the maintenance which should be
performed by the user. By following the general connector care practices
outlined below, the life of the connector can be greatly extended.
• Connectors should be properly torqued as shown below.
• Always tighten or loosen a connector by rotating only the nut. Never
rotate the noise source body.
Connector Type
Torque Specifications
N
1.3 N m (l2 in.-lb.)
APC-7
1.3 N m (12 in.-lb.)
SMA to APC-3.5
0.9 N m (8 in.-lb.)
APC-3.5 to APC-3.5
0.9 N m (8 in.-lb.)
Содержание 346A
Страница 8: ...1 1 General Information ...
Страница 35: ...28 Chapter1 General Information Replaceable Parts Figure 1 10 Preparing for Parts Removal Replacement ...
Страница 41: ...35 3 Using the ENR Data Diskette ...
Страница 48: ...43 4 ENR File Format ...