![Agilent Technologies 34134A Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/agilent-technologies/34134a/34134a_user-manual_2867851006.webp)
6
34134A AC/DC DMM Current Probe
Specifications and characteristics
Safety Specifications
This symbol signifies that the 34134A AC/DC DMM Current Probe is
protected by double or reinforced insulation. Only use specified replacement
parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user
manual before using the instrument.
WARNING
SHOCK HAZARD!
•
Connect the probe to the DMM or voltage measuring instrument before clamping
the probe around a conductor.
•
Never use the probe on circuits rated higher than 660 V.
•
Never leave the probe clamped around a conductor while not connected to a voltage
measuring instrument or DMM.
•
Carefully center the conductor inside the probe jaws and make sure that the probe
is perpendicular to the conductor before closing the jaws.
•
Avoid, if possible, the proximity of other conductors which may create noise.
•
Check the magnetic mating surfaces of the probe jaws; these should be free of dirt,
rust, or other foreign matter.
•
Do not use a probe which is cracked, damaged or has defective leads.
Working voltage
600 V CAT III
Case protection
IP20 per IEC 529
Drop test
1m according to IEC 68-2-32
Mechanical shock
100 G: test per IEC 68-2-27
Vibration
Test per IEC 68-2-6 frequency range 10 Hz to 55 Hz,
amplitude: 0.15mm
!
!
34134A.book Page 6 Wednesday, March 25, 2009 4:09 PM