2
Open:
Compensates leakage current
and capacitance in test fixtures
and cables.
Verifies reliability of
capacitor test connections.
Stores and retrieves up to
10 measurement states.
Selects values for high,
in, and low testing.
Measurement parameter:
Selects the right parameter
for the measurement.
Display:
LCD with back-light. Displays
measurement values, instrument
states, and comparator results.
Test sequence program:
Controls a series of resistance
measurements.
(Charge-measure-discharge)
Agilent 4349B 4-channel high resistance meter
Key parameters and specifications
Agilent 4339B
Agilent 4349B
Test channels
1
2ch, Option 4349B-001
4ch, Option 4349B-700
Test voltage(Vdc)
0.1 to 1000
Requires external power source
1
Measurement parameters
R, I, pv, ps
R, I
Measurement range
(Ω)
10
3
to 1.6x10
16
10
3
to 10
15
Basic accuracy
0.6% 2%
Display resolution
3 / 4 / 5 digits
3 / 4 / 5 digits
Measurement time
10 ms/30 ms/390 ms
9.5 ms/28 ms/98 ms/397 ms
1. External power source required for resistance measurements. Recommendation for external power source for
measurement for 1 G
Ω
sample at 100 Vdc with accuracy = ±10%:
Ripple:
≤
1 mVrms (50/60 Hz)
Wideband noise:
≤
5 µVrms/Hz (50 Hz)
Switching noise:
≤
50 mVrms (100 kHz)