60
16900A, 16902A, and 16903A Service Guide
6
Replacing Assemblies
To remove and replace the I/O board
16900A and 16902A I/O board
1
Perform previous procedures:
•
“To prepare the instrument for disassembly"
•
“To remove and replace the cover"
•
“To remove and replace the CPU tray"
2
Disconnect the front panel reset cable.
3
Using a deep- well 13/16- inch nut driver, remove the hex
nuts that secure the BNC connectors to the rear panel.
4
Using a Pozidrive screwdriver, remove the 4 screws
securing the multiframe input/output connectors to the
rear panel.
5
Using a Torx T10 screwdriver, remove the screw securing
the I/O board to the CPU tray.
6
Reverse this procedure to install the I/O board.
T-10 Screw (1)
I/O Board
Multiframe
Connectors
16900/02A
Rear Panel
Pozidrive
Screws (4)
Hex Nuts (4)
Front Panel
Reset Cable