Agilent Technologies 16092A Скачать руководство пользователя страница 6

6

 

Содержание 16092A

Страница 1: ...t 16092A Test Fixture Operation and Service Manual Third Edition Agilent Part No 16092 90010 February 2000 Printed in Japan Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Establishe...

Страница 2: ...l changes are incorporated December 1986 First Edition part number 16092 90000 November 1998 Second Edition part number 16092 90010 February 2000 Third Edition part number 16092 90010 Safety Summary T...

Страница 3: ...ized modifications to the instrument Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained Dangerous Procedure W...

Страница 4: ...d corresponding to the individual warranty periods of its component products Instruments are warranted for a period of one year Fixtures and adapters are warranted for a period of 90 days During the w...

Страница 5: ...l damages whether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products Fo...

Страница 6: ...6...

Страница 7: ...Contents 7 1 Operation Description 11 2 Service Maintenance 14...

Страница 8: ...8 Contents...

Страница 9: ...ure 1 1 was specially designed for use with the 4191A RF Impedance Analyzer How to use the 16092A performance characteristic data and operating instructions are given in the 4191A s operation and serv...

Страница 10: ...10 Chapter1 Operation Figure 1 1 16092A Test Fixture...

Страница 11: ...m shorting configuration between the High terminal and ground of the test fixture When the test fixture is short circuited the residual resistance and inductance of the fixture terminal can be measure...

Страница 12: ...12 Chapter1 Operation Description...

Страница 13: ...13 2 Service...

Страница 14: ...d view Do not disassemble any further than shown in the figure To order the necessary replacement parts identify them by their Agilent Technologies part number If a defective part is located in an ass...

Страница 15: ...Chapter 2 15 Service Maintenance Figure 2 1 Parts Identification...

Страница 16: ...460 0350 1 SPRING 10 2190 0125 1 WASHER 11 0520 0174 1 SCREW 12 0520 0133 2 SCREW 13 2190 0103 2 WASHER 14 16092 40010 2 HOLDER 15 2190 0125 2 WASHER 16 16092 24012 2 SCREW 17 1460 0352 2 SPRING 18 16...

Страница 17: ...the contact from the connector The replacement procedure is outlined below 1 Grip the Pin vise and fully open the vise chuck by pressing the button on the vise 2 Place the Pin vice over the center con...

Страница 18: ...surfaces it is recommended that the operator perform periodic cleaning as necessary Use a lint free cloth and if a cleaning fluid is needed use isopropylalcohol CAUTION Do not use aromatic or chlorina...