![Agilent Technologies 16047E Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/agilent-technologies/16047e/16047e_operation-and-service-manual_2868563028.webp)
24
Chapter 3
Operation
Measuring 3-terminal device
DUT.
Similarly, connect the lead No. 2 to the guard terminal in order to measure only
characteristics of Z
2
eliminating any effects from Z
1
and Z
3
.
NOTE
Connect your device to the guard terminal with a shortest possible lead. Longer
lead will degrade the guard effect resulting in less accurate measurement result.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com