Agilent InfinityLab LC Series RID User Manual
97
7
Error Information
Refractive Index Detector Specific Error Messages
Purge Valve Fuse Blown
Error ID: 2686
The electronic fuse on the purge valve has been activated.
Recycle Valve Fuse Blown
Error ID: 2687
The electronic fuse on the recycle valve has been activated.
Probable cause
Suggested actions
1
Short in purge valve circuit.
Power cycle the module.
2
Defective purge valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Short in recycle valve circuit.
Power cycle the module.
2
Defective recycle valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.