9
Characteristics and Specifications
Measurement Characteristics
See note 1 below.
Probe Bandwidth
DC to 50 MHz (-3 dB)
Accuracy (Probe Only)*
±1% of reading ±1mV
(DC or 45 Hz to 66 Hz)
Risetime
2
~
7ns
Maximum current
With InfiniiVision Series:
15A peak(DC+ACpeak), 15 ADC, 10 Arms continuous
Max. peak: 30A peak non-continuous (when two
probes are used)
30A peak(DC+ACpeak), 30 ADC, 24 Arms continuous
Max. peak: 32A peak non-continuous (when one
probe is used)
With Infiniium Series:
15A peak, 15 ADC, 10 Arms continuous, 30A peak
non-continuous (per each channel)
(Refer to frequency derating curve, Fig. 2)
Maximum peak current
50A (for Pulse Widths
~
10µs)
Probe sensitivity
0.1V/A
Noise
~
2.5mArms (with 20 MHz Bandwidth limiting)
Insertion impedance
Refer to Fig. 3
Temperature coefficient
~
±2% (0
%
C to 40
%
C)
Effect of external magnetic fields
Equivalent to
~
20mA (for a 400A/m magnetic field,
DC to 60 Hz)
Maximum circuit voltage
300V Cat I (refer to safety considerations and product
compliances)
*This is a specification and is guaranteed at 23
%
C ±3
%
C (73
%
F ±5
%
F).
1. Requires 1M
W
termination. Valid for 23
%
C ±3
%
C (73
%
F ±5
%
F), at least 30 minutes after power on.