36
1141A and 1142A User’s Guide
2
Calibrating the Probe
Bandwidth
Bandwidth
This test checks the high- frequency response of the 1141A
Differential Probe. The bandwidth of the oscilloscope is
characterized first so it is not a factor in the measurement. This
test depends on the accuracy of the termination on the test board
and the termination in the oscilloscope. Both should be with 1%.
Specification (–3 dB, dc coupled): . . . . . . . . . . . . . . . . . . . . . . . . . . . .dc to 200 MHz
1
With the N cable and N- to- BNC adapter, connect the signal
generator to the oscilloscope channel 1 input.
2
Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms).
3
Set the 1142A front panel switches to Local and Zero offset.
4
On the oscilloscope press
AUTOSCALE
, then set the parameters listed
in
Table 7
Equipment Required
Equipment Required Critical Specifications
Recommended
Model/Part
Oscilloscope
400 MHz bandwidth
54830A
Signal Generator
200 MHz at
≈
230 mVrms
8648A
Test Board
No substitute
01141-66504
Cable
Type N (m) 24-inch
11500B
Adapter
Type N (f) to BNC (m)
1250-0077
Table 8
Oscilloscope Settings
Menu
Selection
Setting
TIMEBASE
(time/div)
2 ns/div
CHAN 1
(sensitivity)
(input R)
100 mV/div
50
Ω
DC
ACQUISITION
Sampling Mode
Memory Depth
Sample Rate
Averaging
# of avg
Real Time
Automatic
Automatic
Enabled
32
Содержание 1141A
Страница 4: ...4 1141A and 1142A User s Guide Contents...
Страница 58: ...58 1141A and 1142A User s Guide 3 Adjusting the Probe Attenuator Adapter Adjustment...
Страница 66: ...66 1141A and 1142A User s Guide 4 Service Troubleshooting Figure 38 1142A Probe Control and Power Module...
Страница 76: ...76 1141A and 1142A User s Guide 4 Service Replaceable Parts Figure 44 Cabling Diagram...
Страница 78: ...78 1141A and 1142A User s Guide 4 Service Replaceable Parts Figure 46 1142A Probe Control and Power Module...
Страница 86: ...86 1141A and 1142A User s Guide 4 Service Replaceable Parts...