24
1141A and 1142A User’s Guide
1
Operating the Probe
Remote Operation
Variable Offset
The remote variable offset can be used when the dc couple function
is remotely selected. The offset voltage must be referenced to the
External offset common (pin 4) of the remote input connector. It
must be electrically isolated from the controlling system. The
following table shows the offset range and remote offset
requirements for probe and adapter combinations.
Table 4
Remote Functions Select Truth Table
Remote Input
Function
A1R (Pin 1)
A0R (Pin 6)
0
0
0.05 Hz Reject
0
1
0.5 Hz Reject
1
0
5 Hz Reject
1
1
DC Couple
Table 5
Remote Offset Input Requirements
Offset Range
Remote
Requirements
Probe alone
± 20V
± 10V
Probe with 10x adapter
± 200V
± 10V
Probe with 100x adapter
± 500V
± 2.5V