First Steps with the Fluorescence Detector
2
1100 Series FD Reference Manual
35
2
Set the FLD setpoints according to
(on the local control module
this information is split across separate screens).
3
Start the run.
4
The resulting chromatograms are shown in
5
The excitation maxima is around 250 nm.
Figure 7
FLD Parameters
Figure 8
Biphenyl peak with different excitation wavelengths
In this example
additional excitation
wavelengths (B, C,
D) are used. This will
increase the scan
time and may lower
the performance.
Ex=230 nm
Ex=246 nm
Ex=250 nm
Ex=290 nm
Biphenyl peak