190
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Exchanging the Power Supply
When required
If defective
Tools required
Screwdriver POZI 1 PT3
Wrench 1/4 inch
Wrench 5 mm
Wrench 7 mm
Parts required
Power supply 0950-2528
N O T E
The repair of the power supply assembly comprises exchanging the complete assembly. No
serviceable parts are inside.
Preparations for this procedure:
•
Turn off the lamp.
•
Switch off the detector, and disconnect the cables.
•
Remove the detector from the stack and place it on the workbench.
•
Remove the front cover and top cover, see
•
Remove the processor board, see
“Exchanging the Processor Board"
•
•
Remove the optical unit, see
Содержание 1100 Series