Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
163
Exchanging the Fan
When required
If the fan is defective or noisy
Tools required
Screwdriver POZI 1 PT3
Parts required
Fan assembly 3160-1017
C A U T I O N
The fan must be installed in the correct orientation to ensure optimum cooling and
operation of the detector.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
1
Disconnect the emission and the excitation cables and
from the detector main board.
A
B
A
B
A
EM
EX